FPGA-based digitizer for BGO-based time-of-flight PET

Daehee Lee,Sun Il Kwon
2024-10-12
Abstract:We present a novel FPGA-based bismuth germanate (BGO) time-of-flight (TOF) digitizer, implemented on an FPGA (VC707 evaluation kit, Xilinx). This digitizer was designed to address the recently highlighted characteristics of BGO, which generates scintillation and prompt Cerenkov photons when 511-keV photon interacts with BGO. The developed digitizer independently processes these two types of photons for precise energy and timing measurements. The digitizer incorporates a noise-resistant binary counter that measures energy signals using a time-over-threshold (TOT) method. For timing measurement, we employ an embedded dual-side monitoring time-to-digital converter (TDC), which efficiently captures timing information while maintaining low resource usage. We validated the efficacy of our FPGA-based TOF digitizer through extensive experiments, including both an electrical setup and a coincidence test using BGO pixels. Our evaluations of TOT energy and timing performance utilized two $3 \times 3 \times 20 \ \text{mm}^3$ BGO pixels coupled to CHK-HD MT silicon photomultipliers (SiPMs). The digitizer achieved a coincidence timing resolution (CTR) of 407 ps full width at half maximum (FWHM) for coincidence events falling within the full width at tenth maximum (FWTM) of the measured TOT energy spectrum. Notably, when measured with an oscilloscope, the same detector pair exhibited a CTR of 403 ps FWHM, confirming that the performance of the developed digitizer is comparable to that of oscilloscopes. With the low resource usage of our design, it offers significant potential for scalability, making it particularly promising for multi-channel BGO-based PET systems.
Instrumentation and Detectors,Medical Physics
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to develop a new FPGA (Field - Programmable Gate Array) digitizer for a Time - of - Flight Positron Emission Tomography (TOF PET) system for BGO (BiGeO, Bismuth Germanium Oxide) materials. Specifically, the researchers aim to improve the accuracy of energy and time measurements by independently processing the scintillation photons and Cherenkov photons generated when BGO crystals interact with 511 keV photons. This technological improvement is of great significance for improving the performance of TOF PET systems, especially in terms of cost - effectiveness, because BGO materials are much cheaper than the commonly used LSO (Lutetium Oxyorthosilicate) or LYSO (Lutetium - Yttrium Oxyorthosilicate) materials. ### Main Problems and Solutions 1. **Problem**: Traditional TOF PET Data Acquisition Systems (DAQs) are mainly designed for high - light - yield scintillators such as LSO and LYSO, and these systems have difficulty effectively processing the weaker signals generated in BGO, especially the Cherenkov photon signals used for time measurement. 2. **Solutions**: - **Energy Measurement**: Adopt the Time - over - Threshold (TOT) method, combined with the Noise - Resistant Binary Counter (NRBC), to achieve accurate measurement of BGO energy signals. NRBC filters out noise caused by local spikes through Double - Check Logic (DCL), thereby avoiding incorrect measurements or premature termination of counting operations. - **Time Measurement**: Use an Embedded Dual - Side Monitoring Time - to - Digital Converter (DSM TDC), by simultaneously monitoring the Start of Propagation (SOP) and End of Propagation (EOP) of signal propagation, to efficiently capture time information while maintaining low resource consumption. This method can maintain stable time - measurement characteristics under Power, Voltage, and Temperature (PVT) variation conditions without the need for additional correction circuits. ### Experimental Verification The researchers verified the effectiveness of the proposed FPGA - based TOF digitizer through a series of experiments, including electrical tests and coincidence tests based on BGO pixels. The experimental results show that this digitizer can achieve a Coincidence Timing Resolution (CTR) comparable to that of an oscilloscope. In particular, under the best conditions (such as a 49 V bias voltage and an energy threshold of 250 mV), the CTR reaches 407 ps FWHM (Full Width at Half Maximum), which is very close to the 403 ps FWHM measured using an oscilloscope. ### Conclusion This research has successfully developed an FPGA - based TOF digitizer specifically designed for BGO materials. It not only improves the accuracy of energy and time measurements but also has good scalability due to its low - resource - consumption design, and is especially suitable for multi - channel BGO - based TOF PET systems. This provides new possibilities for the development of low - cost, high - performance TOF PET detectors.