A dosimetric and robustness analysis of Proton Arc Therapy (PAT) with Early Energy Layer and Spot Assignment (ELSA) for lung cancer versus conventional Intensity modulated therapy (IMPT)

Macarena S. Chocan,Sophie Wuyckens,Damien Dasnoy,Dario Di Perri,Elena Borderias Villarruel,Erik Engwall,John A. Lee,Ana M. Barragán-Montero,Edmond Sterpin
2024-09-25
Abstract:Background and purpose: IMPT faces challenges in lung cancer treatment, like maintaining plan robustness for moving tumors against setup, range errors, and interplay effects. Proton Arc Therapy (PAT) is an alternative to maintain target coverage, potentially improving organ at risk (OAR) sparing, reducing beam delivery time (BDT), and enhancing patient experience. We aim to perform a systematic plan comparison study between IMPT and ELSA-PAT to assess its potential for lung cancer treatment. Material and Methods: 14 Lung ELSA-PAT plans were compared retrospectively with IMPT plans. 4D worst-case minimax robust optimization was performed, including 84 scenarios (3%,3 mm). Dosimetry assessment included target (CTV) and important OARs, on nominal and worst-case scenarios. Most relevant normal tissue complication probabilities (NTCP), target coverage robustness against interplay effect and beam delivery time (BDT) were evaluated. Results: CTV D95% and D98% showed no significant difference in comparison. PAT demonstrated better conformality by 66% (p = 0.00012) but delivered a higher heart mean dose (HMD,23%). There was a 2% increase in NTCP 2-year mortality risk with PAT. Total BDT was comparable among techniques. IMPT was more robust than PAT against interplay effect, considering both D1% (1,0 $\pm$ 0.8 Gy vs 1.1 $\pm$ 1.4 Gy) and D98% bandwidths (0.9$\pm$0.9 Gy vs 1.1 $\pm$ 1.3 Gy). Interpretation: both techniques provide a similar level of dose coverage to the target volume. Although PAT improved dose conformality, higher HMD translated into increased heart toxicity, presumably due to chosen planning methodology and OAR proximity to target. Increased energy layers and spots raised PAT beam delivery time, although it could improve daily treatment workflow.
Medical Physics
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to compare the dosimetric differences and planning robustness between proton arc therapy (PAT) based on the early - layer and spot - allocation algorithm (ELSA) and traditional intensity - modulated proton therapy (IMPT) in the treatment of lung cancer. Specifically, the research aims to: 1. **Quality of dose distribution**: Evaluate the performance of PAT and IMPT in terms of target coverage, dose uniformity, and organ - at - risk (OAR) protection. 2. **Planning robustness**: Analyze the robustness of PAT and IMPT when facing the interaction effects caused by setup errors, range uncertainties, and respiratory motion. 3. **Normal - tissue complication probability (NTCP)**: Calculate and compare the possible long - term toxic effects of the two techniques on normal tissues (such as the lung, esophagus, and heart). 4. **Beam - delivery time (BDT)**: Evaluate the beam - delivery time of PAT and IMPT in actual clinical applications and explore whether PAT can simplify the treatment process and reduce the treatment time. Through a comprehensive analysis of these aspects, the study hopes to provide clinicians with comprehensive information on the advantages and disadvantages of PAT and IMPT in the treatment of lung cancer to support more rational treatment decisions.