Spatial resolution studies using point spread function extraction in optically read out Micromegas and GEM detectors

A. Cools,E. Ferrer-Ribas,T. Papaevangelou,E.C. Pollacco,M. Lisowska,F.M. Brunbauer,E. Oliveri,F.J. Iguaz
2024-07-22
Abstract:Optically read out gaseous detectors are used in track reconstruction and imaging applications requiring high granularity images. Among resolution-determining factors, the amplification stage plays a crucial role and optimisations of detector geometry are pursued to maximise spatial resolution. To compare MicroPattern Gaseous Detector (MPGD) technologies, focused low-energy X-ray beams at the SOLEIL synchrotron facility were used to record and extract point spread function widths with Micromegas and GEM detectors. Point spread function width of $\approx$108\,\microns for Micromegas and $\approx$127\,\microns for GEM foils were extracted. The scanning of the beam with different intensities, energies and across the detector active region can be used to quantify resolution-limiting factors and improve imaging detectors using MPGD amplification stages.
Instrumentation and Detectors
What problem does this paper attempt to address?