Cosserat Rods for Modeling Tendon-Driven Robotic Catheter Systems

Pierre-Frédéric Villard,Thomas M. Waite,Robert D. Howe
2024-09-25
Abstract:Tendon-driven robotic catheters are capable of precise execution of minimally invasive cardiac procedures including ablations and imaging. These procedures require accurate mathematical models of not only the catheter and tendons but also their interactions with surrounding tissue and vasculature in order to control the robot path and interaction. This paper presents a mechanical model of a tendon-driven robotic catheter system based on Cosserat rods and integrated with a stable, implicit Euler scheme. We implement the Cosserat rod as a model for a simple catheter centerline and validate its physical accuracy against a large deformation analytical model and experimental data. The catheter model is then supplemented by adding a second Cosserat rod to model a single tendon, using penalty forces to define the constraints of the tendon-catheter system. All the model parameters are defined by the catheter properties established by the design. The combined model is validated against experimental data to confirm its physical accuracy. This model represents a new contribution to the field of robotic catheter modeling in which both the tendons and catheter are modeled by mechanical Cosserat rods and fully-validated against experimental data in the case of the single rod system.
Robotics
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to develop a mechanical model that can accurately simulate the tendon - driven robotic catheter system. Specifically, this model needs to be able to accurately describe the interaction between the catheter and the tendon, as well as their interaction with the surrounding tissues and blood vessels, in order to better control the robotic path and operation. To achieve this goal, the author proposes a mechanical model based on the Cosserat rod theory and combines it with a stable implicit Euler scheme. This model not only considers a simple model of the catheter centerline, but also simulates a single tendon by adding a second Cosserat rod and uses penalty forces to define the constraints in the tendon - catheter system. All model parameters are defined by the catheter properties determined by the design. Finally, the physical accuracy of this model is verified by experimental data, especially in the case of a single - rod system, achieving simultaneous modeling and comprehensive verification of the tendon and the catheter. This marks a new contribution in the field of robotic catheter modeling, that is, for the first time, simultaneously using the mechanical Cosserat rod model to model the tendon and the catheter and comprehensively verifying them through experimental data.