The Virgo Newtonian calibration system for the O4 observing run

Florian Aubin,Eddy Dangelser,Dimitri Estevez,Alain Masserot,Benoît Mours,Thierry Pradier,Antoine Syx,Pierre Van Hove
DOI: https://doi.org/10.1088/1361-6382/ad869c
2024-10-15
Abstract:After initial tests performed during previous observing runs, a Newtonian Calibrator (NCal) system was developed and installed on the Virgo gravitational wave detector for the O4 observing run. This system, which is continuously operated, provides the absolute calibration of Virgo for this run. Its $1\mbox{-}\sigma$ uncertainty of $0.17\%$ on the amplitudes of the injected signals is better than that obtained with other calibration techniques like the photon calibrator (PCal). This paper presents this NCal system and details the different sources of uncertainties
General Relativity and Quantum Cosmology
What problem does this paper attempt to address?