In vivo fundus imaging and computational refocusing with a diffuser-based fundus camera

Corey Simmerer,Marisa Morakis,Lei Tian,Lia Gomez-Perez,T.Y. Alvin Liu,Nicholas J. Durr
2024-06-11
Abstract:Access to eye care can be expanded with high-throughput, easy-to-use, and portable diagnostic tools. Phase mask encoded imaging could improve these aspects of the fundus camera by enabling computational refocusing without any moving parts. This approach circumvents the need to adjust lenses to compensate for refractive errors. We developed a computational fundus camera by introducing a holographic diffuser at the plane conjugate to the ocular pupil, resulting in a laterally shift-invariant point spread function. We demonstrate computational refocusing of a model eye fundus over a large range of defocus errors. We also show computationally refocused, color, in vivo, human fundus images with a $\geq$35-degree field-of-view (FOV). This technology could eventually be combined with the wavefront-sensing capabilities of phase mask encoded imaging to create a compact ophthalmic imaging system that simultaneously captures a fundus image and performs aberrometry.
Optics,Medical Physics
What problem does this paper attempt to address?
This paper proposes a solution to the focusing difficulty problem of traditional fundus cameras during fundus examination. The research team has developed a computational fundus camera based on a diffraction grating, which uses phase mask encoding imaging technology to achieve computational focusing without the need for moving parts. This method bypasses the need to adjust the lens due to refractive errors. In the paper, the researchers placed a holographic diffraction grating on the pupil conjugate plane of the eye to obtain a laterally shift-invariant point spread function, which allows for the calculation of focal errors over a large range for the model eye fundus. They also demonstrated the computational focusing results of in vivo human eye fundus images with at least a 35-degree field of view after chromatic aberration correction. Although portable fundus cameras exist, they often have limited field of view and difficulties with focusing, resulting in low clinical adoption. The proposed computational fundus photography technique has the potential to be combined with wavefront sensing to create a compact optical imaging system that can capture fundus images and perform aberration measurement, increasing the accessibility of fundus examination, especially in resource-limited areas. The research team experimentally verified the optical performance of this system and demonstrated computational focusing images of in vivo fundus, proving its focusing capability under different levels of defocus errors. However, the resolution of this method is lower than that of traditional lens cameras, and future work is needed to improve the resolution to ensure diagnostic capability.