Calibration of the Cryogenic Measurement System of a Resonant Haloscope Cavity

Dong He,Jie Fan,Xin Gao,Yu Gao,Nick Houston,Zhongqing Ji,Yirong Jin,Chuang Li,Jinmian Li,Tianjun Li,Shi-hang Liu,Jia-Shu Niu,Zhihui Peng,Liang Sun,Zheng Sun,Jia Wang,Puxian Wei,Lina Wu,Zhongchen Xiang,Qiaoli Yang,Chi Zhang,Wenxing Zhang,Xin Zhang,Dongning Zheng,Ruifeng Zheng,Jian-yong Zhou
2024-04-16
Abstract:Possible light bosonic dark matter interactions with the Standard Model photon have been searched by microwave resonant cavities. In this paper, we demonstrate the cryogenic readout system calibration of a 7.138 GHz copper cavity with a loaded quality factor $Q_l=10^4$, operated at 22 mK temperature based on a dilution refrigerator. Our readout system consists of High Electron Mobility Transistors as cryogenic amplifiers at 4 K, plus room-temperature amplifiers and a spectrum analyzer for signal power detection. We test the system with a superconducting two-level system as a single-photon source in the microwave frequency regime and report an overall 95.6 dB system gain and -71.4 dB attenuation in the cavity's input channel. The effective noise temperature of the measurement system is 7.5 K.
High Energy Physics - Experiment,High Energy Physics - Phenomenology,Quantum Physics
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to calibrate the 7.138 - GHz copper - cavity resonant haloscope system in the low - temperature measurement system. Specifically, the researchers designed a low - temperature read - out system based on a dilution refrigerator and carried out detailed calibration on it. The purpose of this system is to detect the interaction between light - mass boson dark matter and standard - model photons, especially through a microwave resonant cavity to achieve this goal. In the experiment, the researchers used high - electron - mobility transistors (HEMTs) as low - temperature amplifiers at 4 K, together with amplifiers at room temperature and a spectrum analyzer to detect signal power. To test the system, they used a superconducting two - stage system as a single - photon source and reported a total system gain of 95.6 dB and an attenuation of - 71.4 dB in the cavity input channel. The effective noise temperature of the measurement system is 7.5 K. The key to the paper lies in how to accurately calibrate the attenuation, gain and noise level of the low - temperature measurement system, especially in the wide - band frequency range. This is not only crucial for improving the sensitivity of detecting light - mass boson dark matter, but also provides a technical reference for similar experiments. By using the Mollow triplet of the superconducting two - stage system for calibration, the researchers were able to achieve high - precision measurement of system parameters, which is of great significance for subsequent experimental design and data analysis.