Search for Axion dark matter with the QUAX-LNF tunable haloscope

A. Rettaroli,D. Alesini,D. Babusci,C. Braggio,G. Carugno,D. D'Agostino,A. D'Elia,D. Di Gioacchino,R. Di Vora,P. Falferi,U. Gambardella,A. Gardikiotis,C. Gatti,G. Iannone,C. Ligi,A. Lombardi,G. Maccarrone,A. Ortolan,G. Ruoso,S. Tocci,G. Vidali
2024-05-15
Abstract:We report the first experimental results obtained with the new haloscope of the QUAX experiment located at Laboratori Nazionali di Frascati of INFN (LNF). The haloscope is composed of a OFHC Cu resonant cavity cooled down to about 30 mK and immersed in a magnetic field of 8 T. The cavity frequency was varied in a 6 MHz range between 8.831496 and 8.83803 GHz. This corresponds to a previously unprobed mass range between 36.52413 and 36.5511 $\mu$eV. We don't observe any excess in the power spectrum and set limits on the axion-photon coupling in this mass range down to $g_{a\gamma\gamma} < 0.861 \times 10^{-13}$ GeV$^{-1}$ with the confidence level set at $90\%$.
Instrumentation and Detectors,Cosmology and Nongalactic Astrophysics,High Energy Physics - Experiment
What problem does this paper attempt to address?