Ultra-High precision Compton polarimetry at 2 GeV

A. Zec,S. Premathilake,J. C. Cornejo,M. M. Dalton,C. Gal,D. Gaskell,M. Gericke,I. Halilovic,H. Liu,J. Mammei,R. Michaels,C. Palatchi,J. Pan,K. D. Paschke,B. Quinn,J. Zhang
DOI: https://doi.org/10.1103/PhysRevC.109.024323
2024-02-26
Abstract:We report a high precision measurement of electron beam polarization using Compton polarimetry. The measurement was made in experimental Hall A at Jefferson Lab during the CREX experiment in 2020. A total uncertainty of dP/P=0.36% was achieved detecting the back-scattered photons from the Compton scattering process. This is the highest accuracy in a measurement of electron beam polarization using Compton scattering ever reported, surpassing the ground-breaking measurement from the SLAC Large Detector (SLD) Compton polarimeter. Such uncertainty reaches the level required for the future flagship measurements to be made by the MOLLER and SoLID experiments.
Instrumentation and Detectors,Nuclear Experiment