An experimental system for the study of ultrasound exposure of isolated blood vessels

Anna Tokarczyk,Ian Rivens,E van Bavel,Richard Symonds-Tayler,Gail ter Haar
DOI: https://doi.org/10.1088/0031-9155/58/7/2281
2013-04-07
Abstract:An experimental system designed for the study of the effects of diagnostic or therapeutic ultrasound exposure on isolated blood vessels in the presence or absence of intraluminal contrast agent is described. The system comprised several components. A microscope was used to monitor vessel size (and thus vessel functionality), and potential leakage of intraluminal 70 kDa FITC-dextran fluorescence marker. A vessel chamber allowed the mounting of an isolated vessel whilst maintaining its viability, with pressure regulation for the control of intraluminal pressure and induction of flow for the infusion of contrast microbubbles. A fibre-optic hydrophone sensor mounted on the vessel chamber using a micromanipulator allowed pre-exposure targeting of the vessel to within 150 µm, and monitoring of acoustic cavitation emissions during exposures. Acoustic cavitation was also detected using changes in the ultrasound drive voltage and by detection of audible emissions using a submerged microphone. The suitability of this system for studying effects in the isolated vessel model has been demonstrated using a pilot study of 6 sham exposed and 18 high intensity focused ultrasound exposed vessels, with or without intraluminal contrast agent (SonoVue) within the vessels.