D B A Tran,Olivier Lopez,M Manceau,A Goncharov,Michel Abgrall,H Alvarez-Martinez,R Le Targat,E Cantin,P.-E Pottie,A Amy-Klein,B Darquié
Abstract:We report the development and operation of a frequency-comb-assisted high-resolution mid-infrared molecular spectrometer combining high spectral purity, SI-traceability, wide tunability and high sensitivity. An optical frequency comb is used to transfer the spectral purity of a SI-traceable 1.54 $\mu$m metrology-grade frequency reference to a 10.3 $\mu$m quantum cascade laser (QCL). The near-infrared reference is operated at the French time/frequency metrology institute, calibrated there to primary frequency standards, and transferred to Laboratoire de Physique des Lasers via the REFIMEVE fiber network. The QCL exhibits a sub-10 --15 frequency stability from 0.1 to 10 s and its frequency is traceable to the SI with a total uncertainty better than 4 x 10 --14 after 1-s averaging time. We have developed the instrumentation allowing comb modes to be continuously tuned over 9 GHz resulting in a QCL of record spectral purity uninterruptedly tunable at the precision of the reference over an unprecedented span of 1.4 GHz. We have used our apparatus to conduct sub-Doppler spectroscopy of methanol in a multi-pass cell, demonstrating state-of-art frequency uncertainties down to the few kilohertz level. We have observed weak intensity resonances unreported so far, resolved subtle doublets never seen before and brought to light discrepancies with the HITRAN database. This demonstrates the potential of our apparatus for probing subtle internal molecular processes, building accurate spectroscopic models of polyatomic molecules of atmospheric or astrophysical interest, and carrying out precise spectroscopic tests of fundamental physics.
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to develop a mid - infrared molecular spectrometer that combines high spectral purity, SI traceability, a wide tuning range, and high sensitivity. Specifically, the researchers used optical frequency comb technology to transfer the spectral purity of a 1.54 μm metrology - grade frequency reference traceable to the International System of Units (SI) to a 10.3 μm quantum - cascade laser (QCL). The keys to this technology are:
1. **Improving spectral resolution**: By transferring the spectral purity of the near - infrared reference signal to the mid - infrared region, the QCL can achieve an unprecedentedly high spectral resolution, which is crucial for analyzing complex molecular structures and dynamic processes.
2. **Expanding the tuning range**: The researchers developed an instrument that enables continuous tuning of the frequency - comb mode within a 9 GHz range, thereby achieving continuous tuning of the QCL within a 1.4 GHz range, which is one of the widest tuning ranges reported so far.
3. **Achieving high - sensitivity measurements**: By using a multi - pass absorption cell for saturated - absorption spectroscopy measurements, the system can detect very weak methanol resonance lines that have not been reported before and can resolve the subtle doublet structure, which has not been observed in previous studies.
4. **Verifying database accuracy**: By comparing with the data in the HITRAN database, the researchers found some inconsistencies, indicating that their instrument has extremely high precision and can be used to verify and improve existing spectral databases.
### Main contributions of the paper
- **Technological breakthrough**: Developed a mid - infrared molecular spectrometer that combines high spectral purity, SI traceability, a wide tuning range, and high sensitivity.
- **Scientific applications**: Demonstrated the potential of this instrument in analyzing complex molecular structures and dynamic processes, especially in atmospheric and astrophysics applications.
- **Data verification**: Discovered inaccuracies in the HITRAN database and provided more accurate methanol spectral data.
### Specific implementation methods
1. **Frequency reference transmission**: Use the 1.54 μm fiber laser from the French Time/Frequency Metrology Institute (LNE - SYRTE) as a frequency reference and transmit it to the laboratory through the REFIMEVE fiber network.
2. **Frequency - comb stabilization**: Use an optical frequency comb to convert the 1.54 μm frequency reference signal to the 10.3 μm QCL, ensuring that the frequency stability of the QCL reaches the order of 10^-15.
3. **Tuning - range expansion**: Through a series of phase - locked loop (PLL) techniques, achieve continuous tuning of the QCL within a 1.4 GHz range.
4. **High - sensitivity measurement**: Use a multi - pass absorption cell for saturated - absorption spectroscopy measurements, detect very weak methanol resonance lines, and resolve the subtle doublet structure.
In summary, this paper solves the comprehensive requirements of high resolution, high sensitivity, and a wide tuning range by developing an advanced mid - infrared molecular spectrometer, providing new tools and techniques for molecular spectroscopy research.