Feasibility evaluation of PET scan-time reduction for diagnosing amyloid-β levels in Alzheimer's disease patients using a deep-learning-based denoising algorithm
Zhao Peng,Ming Ni,Hongming Shan,Yu Lu,Yongzhe Li,Yifan Zhang,Xi Pei,Zhi Chen,Qiang Xie,Shicun Wang,X George Xu,X. George Xu
DOI: https://doi.org/10.1016/j.compbiomed.2021.104919
IF: 7.7
2021-11-01
Computers in Biology and Medicine
Abstract:PurposeTo shorten positron emission tomography (PET) scanning time in diagnosing amyloid-β levels thus increasing the workflow in centers involving Alzheimer's Disease (AD) patients.MethodsPET datasets were collected for 25 patients injected with 18F-AV45 radiopharmaceutical. To generate necessary training data, PET images from both normal-scanning-time (20-minute) as well as so-called "shortened-scanning-time" (1-minute, 2-minute, 5-minute, and 10-minute) were reconstructed for each patient. Building on our earlier work on MCDNet (Monte Carlo Denoising Net) and a new Wasserstein-GAN algorithm, we developed a new denoising model called MCDNet-2 to predict normal-scanning-time PET images from a series of shortened-scanning-time PET images. The quality of the predicted PET images was quantitatively evaluated using objective metrics including normalized-root-mean-square-error (NRMSE), structural similarity (SSIM), and peak signal-to-noise ratio (PSNR). Furthermore, two radiologists performed subjective evaluations including the qualitative evaluation and a five-point grading evaluation. The denoising performance of the proposed MCDNet-2 was finally compared with those of U-Net, MCDNet, and a traditional denoising method called Gaussian Filtering.ResultsThe proposed MCDNet-2 can yield good denoising performance in 5-minute PET images. In the comparison of denoising methods, MCDNet-2 yielded the best performance in the subjective evaluation although it is comparable with MCDNet in objective comparison (NRMSE, PSNR, and SSIM). In the qualitative evaluation amyloid-β positive or negative results, MCDNet-2 were found to achieve a classification accuracy of 100%.ConclusionsThe proposed denoising method has been found to reduce the PET scan time from the normal level of 20 minutes to 5 minutes but still maintaining acceptable image quality in correctly diagnosing amyloid-β levels. These results suggest strongly that deep learning based methods such as ours can be an attractive solution to the clinical needs to improve PET imaging workflow.
engineering, biomedical,computer science, interdisciplinary applications,mathematical & computational biology,biology