A modular table-top setup for ultrafast x-ray diffraction

W. Lu,M. Nicoul,U. Shymanovich,A. Tarasevitch,M. Horn-von Hoegen,D. von der Linde,K. Sokolowski-Tinten
DOI: https://doi.org/10.1063/5.0181132
IF: 1.6
2024-01-01
Review of Scientific Instruments
Abstract:We present a table-top setup for femtosecond time-resolved x-ray diffraction based on a Cu Kα (8.05 keV) laser driven plasma x-ray source. Due to its modular design, it provides high accessibility to its individual components (e.g., x-ray optics and sample environment). The Kα-yield of the source is optimized using a pre-pulse scheme. A magnifying multilayer x-ray mirror with Montel–Helios geometry is used to collect the emitted radiation, resulting in a quasi-collimated flux of more than 105 Cu Kα photons/pulse impinging on the sample under investigation at a repetition rate of 10 Hz. A gas ionization chamber detector is placed right after the x-ray mirror and used for the normalization of the diffraction signals, enabling the measurement of relative signal changes of less than 1% even at the given low repetition rate. Time-resolved diffraction experiments on laser-excited epitaxial Bi films serve as an example to demonstrate the capabilities of the setup. The setup can also be used for Debye–Scherrer type measurements on poly-crystalline samples.
instruments & instrumentation,physics, applied