Sampling and analysis of oxygen in the void volume of ampuls and vials.

L. F. Cullen,G. Papariello
DOI: https://doi.org/10.1002/JPS.2600590118
Journal of Pharmacy and Science
Abstract:A technique recognized as a considerable aid in stabilizing parenteral formulations sensitive to oxidative decomposition consists of replacing air in the container's headspace with an inert atmosphere. To monitor and investigate inert gas flushing techniques and to assist in evaluating antioxidant suitability, a procedure is described for the sampling and analysis of gaseous oxygen content in the void volume headspace of ampuls and vials. The procedure utilizes a paramagnetic oxygen analyzer in conjunction with a unique sampling cell and flow system which permit the collection of the headspace gas without atmospheric contamination. The sampling device and flow system were designed and evaluated to provide the necessary sensitivity, accuracy, and precision for the quantitative analysis of low air contaminant levels in the sample gas. With this method it is possible to analyze nine samples per hour with a relative standard deviation of 3.9% at a 19-μl. air level (i.e., 5 mcg. of oxygen), which corresponds to approximately 2% of the theoretical gaseous void volume of a typical 1-ml. ampul product. Accuracy of the technique is demonstrated by experimentally confirming the relationship of headspace oxygen consumption to intact bisulfite and metabisulfite antioxidant content, following accelerated storage, in sealed ampuls.