Dual-aptamer Drift Cancelling Techniques to Improve Long-term Stability of Real-Time Structure-Switching Aptasensors

Ya-Chen Tsai,Wei-Yang Weng,Yu-Tong Yeh,Jun-Chau Chien
DOI: https://doi.org/10.48550/arXiv.2212.14491
2022-12-30
Abstract:This paper presents a dual-aptamer scheme to cancel the signal drifts from structure-switching aptamers during long-term monitoring. Electrochemical aptamer-based (E-AB) biosensors recently demonstrated their great potential for in vivo continuous monitoring. Nevertheless, the detection accuracy is often limited by the signaling drifts. Conventionally, these drifts are removed by the kinetic differential measurements (KDM) when coupled with square-wave voltammetry. Yet we discover that KDM does not apply to every aptamer as the responses at different SWV frequencies heavily depend on its structure-switching characteristics and the redox reporters' electron transfer (ET) kinetics. To this end, we present a "dual-aptamer" scheme that uses two aptamers responding differentially to the same molecular target for drift cancellation. We identify these paired aptamers through (1) screening from the existing aptamers pool and (2) engineering the signaling behavior of the redox reporters. We demonstrate their differential signaling to ampicillin and ATP molecules and show that the aptamer pair bears common drifts in undilute goat serum. Through cancellation, sensor drift is reduced by 370-fold. Benefiting from the "differential" signaling, the recording throughput is also doubled using differential readout electronics. The authors believe the proposed technique is beneficial for long-term in vivo monitoring.
Biomolecules,Biological Physics
What problem does this paper attempt to address?
The problem that this paper attempts to solve is the decline in detection accuracy caused by signal drift of electro - chemical aptamer - based (E - AB) biosensors in long - term monitoring. Although E - AB biosensors show great potential in continuous in - vivo monitoring, the detection accuracy is often affected by signal drift. The traditional solution is to use the kinetic differential measurement (KDM) technique to remove these drifts by combining square - wave voltammetry (SWV). However, it has been found that KDM is not applicable to all types of aptamers, because the responses at different frequencies strongly depend on their structural conversion characteristics and the electron transfer (ET) kinetics of redox reporter molecules. Therefore, the paper proposes a "dual - aptamer" scheme, that is, using two aptamers that react differently to the same molecular target for drift elimination, thereby improving the stability of long - term monitoring. Specifically, the paper proposes two methods to develop such aptamer pairs: 1. **Screening from existing aptamer libraries**: Look for those aptamers targeting similar molecular groups, especially aptamers for small - molecule antibiotic drugs, such as screening ampicillin and doxorubicin aptamers as a pair. 2. **Changing the position of the redox label**: Modify the position of the redox label on constructs with the same sequence to achieve differential signals upon target binding. For example, use ATP - binding aptamers and connect redox reporter molecules at different positions so that they exhibit different ET kinetics upon target binding. The paper shows the results of 24 - hour measurements in undiluted goat serum using this method, indicating that the drift can be reduced by 370 times. In addition, by using a customized two - channel read - out electronic device to monitor two electrodes simultaneously, the recording throughput is increased. The authors believe that the proposed technology is very beneficial for long - term in - vivo monitoring.