Single-shot quantitative aberration and scattering length measurements in mouse brain tissues using an extended-source Shack-Hartmann wavefront sensor

Sophia Imperato,Fabrice Harms,Antoine Hubert,Mathias Mercier,Laurent Bourdieu,Alexandra Fragola
DOI: https://doi.org/10.1364/OE.456651
2022-03-17
Abstract:Deep fluorescence imaging in mammalian brain tissues remains challenging due to scattering and optical aberration-induced loss in signal and resolution. Correction of aberrations using adaptive optics (AO) requires their reliable measurement in the tissues. Here, we show that an extended-source Shack-Hartmann wavefront sensor (ESSH) allows quantitative aberration measurements through fixed brain slices with a thickness up to four times their scattering length. We demonstrate in particular that this wavefront measurement method based on image correlation is more robust to scattering compared to the standard centroid-based approach. Finally, we obtain a measurement of the tissue scattering length taking advantage of the geometry of a Shack-Hartmann sensor.
Optics,Biological Physics
What problem does this paper attempt to address?
### Problems the Paper Attempts to Solve This paper aims to address the issues of signal loss and resolution degradation caused by scattering and optical aberrations during deep fluorescence imaging in mammalian brain tissue. Specifically, the authors propose a method based on an Extended Source Shack-Hartmann (ESSH) wavefront sensor to quantitatively measure aberrations and scattering length in fixed brain slices. Through this method, the authors hope to achieve more accurate and faster wavefront measurements under high scattering conditions, thereby improving the application of adaptive optics (AO) in optical microscopy. ### Main Research Content 1. **Background and Challenges**: - The main challenges faced in deep fluorescence imaging in mammalian brain tissue are signal loss and resolution degradation due to scattering and optical aberrations. - Adaptive optics (AO) can improve these issues by correcting aberrations, but it requires reliable measurement of aberrations within the tissue. 2. **Methods**: - Use an Extended Source Shack-Hartmann (ESSH) wavefront sensor for wavefront measurement. - Calculate wavefront displacement using an image correlation method instead of the traditional centroid method to improve robustness against scattering. - Measure the scattering length of the tissue using the geometric properties of the Shack-Hartmann sensor. 3. **Experimental Design**: - Implement the ESSH wavefront sensor on a custom-built fluorescence microscope. - Use fixed mouse brain slices of different thicknesses as scattering samples. - Measure the scattering coefficients and spherical aberrations at different wavelengths. 4. **Results**: - The ESSH method demonstrated higher robustness and accuracy under high scattering conditions. - Successfully measured the scattering length of fixed mouse brain tissue and verified its dispersion characteristics at different wavelengths. - Compared to the traditional centroid method, the ESSH method showed better performance under low signal-to-noise ratio conditions. ### Conclusion This study demonstrates the effectiveness of the ESSH wavefront sensor in quantitatively measuring aberrations and scattering length in highly scattering brain tissue. This method enhances the potential application of adaptive optics in deep fluorescence imaging, particularly in mammalian brain tissue. By reducing the impact of scattering on wavefront measurements, the ESSH method is expected to significantly improve the quality and resolution of deep tissue imaging.