Quantitative Evaluation of the Lumbar Multifidus Muscle by Shear Wave Dispersion in Healthy Adults: A Preliminary Study

Qiuxiang Chen,Zhenzhou Li,Meifang Deng,Jiayu Peng,Xiangming Zhu,Chaoxue Zhang
DOI: https://doi.org/10.1002/jum.16416
2024-01-21
Journal of Ultrasound in Medicine
Abstract:Objectives To investigate the application value of shear wave dispersion (SWD) in healthy adults with the lumbar multifidus muscle (LMM), to determine the range of normal reference values, and to analyze the influences of factors on the parameter. Methods Ninety‐five healthy volunteers participated in the study, from whom 2‐dimensional, shear wave elastography (SWE), and SWD images of the bilateral LMM were acquired in three positions (prone, standing, and anterior flexion). Subcutaneous fat thickness (SFH), SWE velocity, and SWD slope were measured accordingly for analyses. Results The mean SWD slope of the bilateral LMM in the prone position was as follows: left: 14.8 ± 3.1 (m/second)/kHz (female) and 13.0 ± 2.5 (m/second)/kHz (male); right: 14.8 ± 3.7 (m/second)/kHz (female) and 14.2 ± 3.4 (m/second)/kHz (male). In the prone position, there was a weak negative correlation between the bilateral LMM SWD slope of activity level 2 and level 1 (β = −1.5 (2 versus 1, left), −1.9 (2 versus 1, right), all P