A novel range telescope concept for proton CT

Marc Granado-González,César Jesús-Valls,Thorsten Lux,Tony Price,Federico Sánchez
DOI: https://doi.org/10.1088/1361-6560/ac4b39
2021-11-27
Abstract:Proton beam therapy can potentially offer improved treatment for cancers of the head and neck and in paediatric patients. There has been a sharp uptake of proton beam therapy in recent years as improved delivery techniques and patient benefits are observed. However, treatments are currently planned using conventional x-ray CT images due to the absence of devices able to perform high quality proton computed tomography (pCT) under realistic clinical conditions. A new plastic-scintillator-based range telescope concept, named ASTRA, is proposed here to measure the proton's energy loss in a pCT system. Simulations conducted using GEANT4 yield an expected energy resolution of 0.7\%. If calorimetric information is used the energy resolution could be further improved to about $0.5\%$. In addition, the ability of ASTRA to track multiple protons simultaneously is presented. Due to its fast components, ASTRA is expected to reach unprecedented data collection rates, similar to $10^8$ protons/s. The performance of ASTRA has also been tested by simulating the imaging of phantoms. The results show excellent image contrast and relative stopping power reconstruction.
Medical Physics,Instrumentation and Detectors
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to develop a new proton CT system in order to improve the planning accuracy of proton beam therapy. Specifically, the paper proposes a new concept of plastic scintillator range telescope named ASTRA for measuring the energy loss of protons in the pCT system. Current proton beam therapy plans rely on traditional X - ray CT images because of the lack of devices capable of performing high - quality proton computed tomography (pCT) under actual clinical conditions. This results in an uncertainty of about 1.6% (mono - energy CT) or 0.7% (dual - energy CT). Directly measuring the relative stopping power (RSP) through pCT can eliminate the systematic error of mapping from photon measurement to RSP, shorten the scanning time, reduce the toxicity rate, and provide pre - treatment image verification under repeated conditions. The ASTRA design proposed in the paper aims to overcome the limitations of existing technologies, such as high production costs, low achievable measurement rates, or insufficient imaging resolution. The design features of ASTRA include: 1. **High data acquisition rate**: Due to its fast components, ASTRA is expected to reach a data acquisition rate of 10^8 protons per second. 2. **Multi - proton simultaneous tracking ability**: ASTRA can simultaneously track multiple protons, which is crucial for improving the overall efficiency of the system and reducing the inefficiency of the detector when it is not possible to perfectly control the delivery of only one proton at a time. 3. **High energy resolution**: Simulation results show that the expected energy resolution of the simulation using GEANT4 is 0.7%, and if calorimetric information is used, the energy resolution can be further improved to about 0.5%. Overall, the goal of this paper is to promote the development of proton CT technology by proposing the innovative design of ASTRA, thereby improving the planning and implementation of proton beam therapy.