Treatment of acute Achilles tendon ruptures with Achillon device: clinical outcomes and kinetic gait analysis

Ignacio Martínez Garrido,Juan Cervera Deval,Marta Navarro Bosch,Daniel Herrero Mediavilla,Vicente Pellicer Garcia,María Sánchez González
DOI: https://doi.org/10.1016/j.fas.2009.10.014
Abstract:Background: We reviewed the outcomes of the Achillon minimally invasive suture system and an early semi-functional rehabilitation program for the treatment of acute Achilles tendon ruptures. Methods: From December 2006 to April 2008, 18 consecutive patients with acute Achilles tendon ruptures were retrospectively reviewed with an average follow-up of 22 months. Clinical data were assessed with the patient satisfaction and the AOFAS hindfoot score. Biomechanical gait parameters were obtained using the NedAMH/IBV dynamometric platform. Results: At last follow-up the AOFAS score was 98 (range, 89-100) and correlated well with kinetic gait analysis. All patients regained normal range of motion and were able to resume their previous activities after six months, with a high rate of satisfaction. Three patients had scar adhesions. There were no cases of recurrent rupture or nerve injury. Conclusions: The Achillon device allows semi-functional rehabilitation and provides satisfactory results with a low rate of complication.
What problem does this paper attempt to address?