Analytical and Clinical Validation of a High Accuracy Fully Automated Digital Immunoassay for Plasma Phospho-Tau 217 for Clinical Use in Detecting Amyloid Pathology

David H. Wilson,Meenakshi Khare,Gallen Triana-Baltzer,Michele Wolfe,Patrick Sheehy,Karen Copeland,Lyndal Hesterberg,Ann-Jeanette Vasko,Wiesje M. van der Flier,Inge M.W. Verberk,Charlotte E. Teunissen,Mike Miller
DOI: https://doi.org/10.1101/2024.10.31.24316186
2024-11-04
Abstract:ABSTRACT Objectives: Phospho-tau 217 (p-Tau 217) in plasma has emerged as a promising biomarker for detecting amyloid pathology in plasma as an aid in diagnosis of Alzheimers disease. Because of its low abundance in plasma, reliable quantitation in blood is challenging. Digital single molecule array (Simoa) technology provides higher analytical sensitivity than possible with conventional analog immunoassays, enabling precise measurement of plasma p-Tau 217. We report the analytical and clinical validation of a Simoa digital immunoassay for p-Tau 217 across highly diverse clinical cohorts that meets consensus criteria for confirmatory test performance to aid the diagnosis of Alzheimers disease in people with cognitive impairment. Methods: A Simoa p-Tau 217 assay utilizing a 2-cutoff approach was analytically validated using industry standard protocols, and diagnostic thresholds were clinically validated across 2 clinically diverse independent cohorts (total n = 873) using amyloid PET or CSF biomarkers as comparators. Results: The assay exhibited acceptable analytical characteristics, including analytical sensitivity enabling measurement of plasma p-Tau 217 in all clinical samples and a lower limit of quantitation of 0.006 pg/ml. For samples not in the intermediate zone between the 2 cutoffs, the assay gave a clinical sensitivity of 90.3%, a specificity of 91.3%, and overall accuracy of 90.7%. 30.9% of the samples fell into the intermediate zone between the lower and upper cutoffs (0.04 and 0.09 pg/mL respectively). With an amyloid prevalence of 50%, typical for mild cognitive impairment, the positive and negative predictive values were 91.4% and 90.4% respectively. Conclusions: The analytical characteristics are suitable for implementation of the Simoa p-Tau 217 assay as a lab developed test under the Clinical Laboratory Improvement Act (CLIA), and the clinical performance characteristics meet consensus criteria for a confirmatory plasma test to aid in Alzheimers diagnosis.
What problem does this paper attempt to address?