A novel method (RIM-Deep) enhances imaging depth and resolution stability of deep-cleared brain tissue in inverted confocal microscopy

Yisi Liu,Pu Wang,Junjie Zou,Hongwei Zhou
DOI: https://doi.org/10.1101/2024.07.19.604108
2024-07-23
Abstract:The increasing use of tissue clearing techniques underscores the urgent need for cost-effective and simplified deep imaging methods. While traditional inverted confocal microscopes excel in high-resolution imaging of tissue sections and cultured cells, they face limitations in deep imaging of cleared tissues due to refractive index mismatches between the immersion media of objectives and sample container. To overcome these challenges, the RIM-Deep was developed to significantly improve deep imaging capabilities without compromising the normal function of the confocal microscope. This system facilitates deep immunofluorescence imaging of the prefrontal cortex in cleared macaque tissue, extending imaging depth from 2 mm to 5 mm. Applied to an intact and cleared Thy1-EGFP mouse brain, the system allowed for clear axonal visualization at high imaging depth. Moreover, this advancement enables large-scale, deep 3D imaging of intact tissues. In principle, this concept can be extended to any imaging modality, including existing inverted wide-field, confocal, and two-photon microscopy. This would significantly upgrade traditional laboratory configurations and facilitate the study of connectomics in the brain and other tissues.
Neuroscience
What problem does this paper attempt to address?
The paper is primarily dedicated to addressing the technical challenges encountered when using inverted confocal microscopy for deep imaging of cleared tissues, particularly brain tissues. Specifically, the research team developed a new method called RIM-Deep, aimed at overcoming the following challenges: 1. **Imaging limitations due to refractive index mismatch**: Traditional inverted confocal microscopes face resolution degradation when performing deep imaging on cleared tissues because of the refractive index mismatch between the objective immersion medium and the sample container. 2. **Limited deep imaging capability**: While inverted confocal microscopes excel at high-resolution imaging of thin sections and cultured cells, they have limitations when it comes to deep imaging of thick, cleared tissues. 3. **Extension of imaging depth**: One of the research team's goals is to significantly increase imaging depth to achieve clear visualization of deep structures in thick tissues such as brain tissue. By designing and implementing the RIM-Deep system, the researchers addressed the aforementioned issues. The RIM-Deep system includes an immersion chamber designed around the objective optical components and a sample holder integrated into the motorized X-Y-Z stage, effectively stabilizing the refractive index between the objective and the sample medium. When used with immersion objectives, the RIM-Deep system enables high-resolution deep imaging, significantly enhancing the capabilities of inverted confocal microscopy. Through experimental validation, the RIM-Deep system successfully increased the imaging depth from 2 millimeters to 5 millimeters while maintaining good resolution stability within this depth range. Additionally, the system was applied to a fully cleared Thy1-EGFP mouse brain, achieving clear axon visualization; and successfully constructed large-scale, deep 3D imaging, which is of great significance for the study of intact tissues. Finally, the system's effectiveness was also demonstrated in the study of vascular damage in mouse brain stroke, showing its great potential in neuroscience research.