A Label-free Optical Biosensor-Based Point-of-Care Test for the Rapid Detection of Monkeypox Virus

Mete Aslan,Elif Seymour,Howard Brickner,Alex Clark,Iris Celebi,Michael B. Townsend,Panayampalli Subbian Satheshkumar,Megan Riley,Aaron Carlin,Selim Unlu,Partha Ray
DOI: https://doi.org/10.1101/2024.07.03.24309903
2024-07-05
Abstract:Diagnostic approaches that combine the high sensitivity and specificity of laboratory-based digital detection with the ease of use and affordability of point-of-care (POC) technologies could revolutionize disease diagnostics. This is especially true in infectious disease diagnostics, where rapid and accurate pathogen detection is critical to curbing the spread of disease. We have pioneered an innovative label-free digital detection platform that utilizes Interferometric Reflectance Imaging Sensor (IRIS) technology. IRIS leverages light interference from an optically transparent thin film, eliminating the need for complex optical resonances to enhance the signal by harnessing light interference and the power of signal averaging in shot-noise-limited operation to achieve virtually unlimited sensitivity. In our latest work, we have further improved our previous 'Single-Particle' IRIS (SP-IRIS) technology by allowing the construction of the optical signature of target nanoparticles (whole virus) from a single image. This new platform, 'Pixel-Diversity' IRIS (PD-IRIS), eliminated the need for z-scan acquisition, required in SP-IRIS, a time-consuming and expensive process, and made our technology more applicable to POC settings. Using PD-IRIS, we quantitatively detected the Monkeypox virus (MPXV), the etiological agent for Monkeypox (Mpox) infection. MPXV was captured by anti-A29 monoclonal antibody (mAb 69-126-3) on Protein G spots on the sensor chips and were detected at a limit-of-detection (LOD) - of 200 PFU/ml (~3.3 attomolar). PD-IRIS was superior to the laboratory-based ELISA (LOD - 1800 PFU/mL) used as a comparator. The specificity of PD-IRIS in MPXV detection was demonstrated using Herpes simplex virus, type 1 (HSV-1), and Cowpox virus (CPXV). This work establishes the effectiveness of PD-IRIS and opens possibilities for its advancement in clinical diagnostics of Mpox at POC. Moreover, PD-IRIS is a modular technology that can be adapted for the multiplex detection of pathogens for which high-affinity ligands are available that can bind their surface antigens to capture them on the sensor surface.
What problem does this paper attempt to address?
The main goal of this paper is to develop a Point-of-Care (POC) technology based on a novel optical biosensor for the rapid and accurate detection of Monkeypox Virus (MPXV). Specifically, the research team has developed a technology called "Pixel Diversity Interferometric Reflectance Imaging Sensor" (PD-IRIS). ### Main Issues Addressed 1. **Improving Detection Speed and Accuracy**: Existing monkeypox detection methods, such as quantitative real-time PCR (qPCR), are accurate but can only be performed in specialized laboratories, making them unsuitable for point-of-care diagnostics. On the other hand, antigen-antibody tests are convenient but have lower sensitivity and specificity. Therefore, there is a need for a detection technology that is both accurate and usable in point-of-care settings. 2. **Simplifying the Operation Process**: Traditional technologies like SP-IRIS require complex optical scanning processes, which increase cost and operational difficulty. PD-IRIS aims to reduce operational complexity and cost by simplifying the detection steps, making it more suitable for point-of-care environments. 3. **Enhancing Sensitivity and Specificity**: By improving the optical signal acquisition method, PD-IRIS achieves higher detection sensitivity (down to 200 PFU/ml, approximately 3.3 attomolar) and demonstrates good specificity in comparative experiments with other viruses (such as Herpes Simplex Virus Type 1 HSV-1 and Cowpox Virus CPXV). ### Technological Innovations - **No Need for z-Scanning**: PD-IRIS eliminates the time-consuming and expensive z-scanning process, instead constructing the optical characteristics of target nanoparticles (whole viruses) in a single image, greatly simplifying the operation process. - **Multi-Wavelength Imaging**: By using multi-parameter optical excitation and collecting reflected waves, PD-IRIS can capture information about target particles in a single snapshot, improving detection efficiency. - **Modular Design**: PD-IRIS is a modular technology that can be adaptively adjusted for different pathogens with high-affinity ligands to capture antigens on the surfaces of these pathogens. In summary, this research aims to address the need for rapid and accurate detection of Monkeypox Virus through PD-IRIS technology and is expected to be widely applicable in point-of-care diagnostic environments.