Towards a Buruli Ulcer Rapid Diagnostic Test that Targets Mycolactone
Marina Siirin,Bijan Pedram,Maria J. Gonzalez-Moa,Louisa Warryn,Rie Yotsu,Jean M. Saunders,Aaron E. Saunders,Richard K. Baldwin,Jessica L. Porter,Timothy P. Stinear,Israel Cruz-Mata,Dziedzom Komi de Souza,Gerd Pluschke,Marco A. Biamonte
DOI: https://doi.org/10.1101/2024.01.23.24301643
2024-01-23
Abstract:We report the development of a prototype rapid diagnostic test for Buruli ulcer, an ulcerative necrotizing skin disease caused by . The test was designed to detect mycolactone, a metabolite unique to . The chief technical challenge was to develop a simple workflow to extract trace amounts of mycolactone from wound exudates collected with a swab and, after a concentration step, to visualize the mycolactone by means of a lateral flow assay. This was achieved by utilizing a mouse monoclonal antibody specific for mycolactone and magnetic gold nanoshells. The latter are a novel class of reporter particles consisting of a ferrite core, a silica gel middle layer that serves to decrease the overall density of the nanoparticle and facilitate its resuspension in aqueous media, and an outer layer of gold, which provides a dark coloration through plasmon resonance effects. These nanoparticles, once conjugated to the anti-mycolactone antibody, enable the immunomagnetic concentration of the targeted analyte and its detection by lateral flow assay. The test procedure can be conducted within 2 hours with a magnetic rack and no powered instrumentation is required. The test can detect as little as 3.5–7 ng of mycolactone collected on a swab.
Infectious Diseases (except HIV/AIDS)
What problem does this paper attempt to address?