Autotuning Plug-and-Play Algorithms for MRI

Saurav K. Shastri,Rizwan Ahmad,Philip Schniter
DOI: https://doi.org/10.48550/arXiv.2012.00887
2020-12-02
Abstract:For magnetic resonance imaging (MRI), recently proposed "plug-and-play" (PnP) image recovery algorithms have shown remarkable performance. These PnP algorithms are similar to traditional iterative algorithms like FISTA, ADMM, or primal-dual splitting (PDS), but differ in that the proximal update is replaced by a call to an application-specific image denoiser, such as BM3D or DnCNN. The fixed-points of PnP algorithms depend upon an algorithmic stepsize parameter, however, which must be tuned for optimal performance. In this work, we propose a fast and robust auto-tuning PnP-PDS algorithm that exploits knowledge of the measurement-noise variance that is available from a pre-scan in MRI. Experimental results show that our algorithm converges very close to genie-tuned performance, and does so significantly faster than existing autotuning approaches.
Information Theory
What problem does this paper attempt to address?