Imaging and Spectral Performance of a 60 μm Pitch CdTe Double-Sided Strip Detector

Kento Furukawa,Shunsaku Nagasawa,Lindsay Glesener,Miho Katsuragawa,Shin'ichiro Takeda,Shin Watanabe,Tadayuki Takahashi
DOI: https://doi.org/10.1016/j.nima.2020.164378
2020-06-11
Abstract:We have evaluated the performance of a fine pitch CdTe Double-sided Strip Detector (CdTe-DSD), which was originally developed for the focal plane detector of a hard X-ray telescope to observe the Sun. The detector has a thickness of 750 um and has 128 strip electrodes with a 60 um strip pitch orthogonally placed on both sides of the detector and covers an energy range 4 keV to 80 keV. The study of the depth of photon interaction and charge sharing effects are of importance in order to provide good spectroscopic and imaging performance. We study the tail structure observed in the spectra caused by charge trapping and develop a new method to reconstruct the spectra based on induced charge information from both anode and cathode strips. By applying this method, energy resolutions (FWHM) of 0.76 keV and 1.0 keV can be obtained at photon energies of 14 keV and 60 keV, respectively, if the energy difference between the anode and cathode is within 1 keV. Furthermore, the tail component at 60 keV is reduced, and the energy resolution of the 60 keV peak is improved from 2.4 keV to 1.5 keV (FWHM) if the energy difference is greater than 1 keV. In order to study the imaging performance, we constructed a simple imaging system using a 5 mm thick tungsten plate that has a pinhole with a diameter of 100 um. We utilize a Ba-133 radioisotope of 1 mm in diameter as a target source in combination with a 100 um slit made from 0.5 mm thickness tungsten. We imaged the Ba-133 source behind the 100 um slit using a 30 keV peak, with a 100 um pinhole placed at the center of the source-detector distance. By applying a charge sharing correction between strips, we have succeeded in obtaining a position resolution better than the strip pitch of 60 um.
Instrumentation and Methods for Astrophysics,High Energy Physics - Experiment,Instrumentation and Detectors
What problem does this paper attempt to address?
The problems that this paper attempts to solve mainly focus on two aspects: 1. **Improving spectral performance**: It is mentioned in the study that due to the charge trapping effect, a low - energy tail structure will appear in the cathode - side spectrum at high - energy peaks. To improve this phenomenon, the researchers developed a new spectral reconstruction method based on the induced charge information on the anode and cathode stripes. Through this method, when the energy difference between the anode and the cathode is within 1 keV, energy resolutions (FWHM) of 0.76 keV and 1.0 keV can be obtained at photon energies of 14 keV and 60 keV respectively. For events with an energy difference greater than 1 keV, the energy resolution of the 60 keV peak is improved from 2.4 keV to 1.5 keV (FWHM). 2. **Improving imaging performance**: The researchers also strive to achieve sub - strip resolution, that is, a resolution finer than the 60 µm stripe width, by considering the charge - sharing information between adjacent channels. To this end, they constructed a simple imaging system, using a 100 µm pinhole on a 5 mm - thick tungsten plate, and using a 133Ba radiation source combined with a 100 µm - wide tungsten slit as the target source. By applying charge - sharing correction, a position resolution better than the 60 µm stripe spacing was successfully obtained. Overall, this study aims to meet the requirements of hard X - ray telescopes for solar observations by improving the spectral and imaging performance of the CdTe double - side - striped detector.