High Spatiotemporal Resolution Contrast-Free Ultrasound Microvascular Imaging Using Adaptive Weight-Based Nonlinear Compounding

Liyuan Jiang,Hanbing Chu,Yang Liu,Jiacheng Liu,Xiao Su,Yichen Yan,Meiling Liang,Yiran Chen,Chaoyang Zhang,Feiqian Wang,Yujin Zong,Mingxi Wan
DOI: https://doi.org/10.1109/tim.2024.3470975
IF: 5.6
2024-10-11
IEEE Transactions on Instrumentation and Measurement
Abstract:Simultaneously improving the spatial and temporal resolution of ultrafast power Doppler imaging (UPDI) is conducive to monitoring rapid changes in microvascular morphology and dynamics and increasing quantification accuracy. Grouped angle singular value decomposition (GA-SVD) has been studied to suppress clutter at enhanced temporal resolution, but it struggles to remove spatially varying noise. This study presents an adaptive weight-based nonlinear compounding method called generalized frame-multiply-and-sum (FMAS) factor based on pth root compression (p-GFMAS) to address the above legacy noise and improve the spatial resolution of UPDI at better temporal resolution. p-GFMAS utilizes angular coherence extracted based on the autocorrelation function to emphasize coherent blood signals and suppress incoherent noise, where the adaptive effect is adjusted by p-value. The effectiveness of the p-GFMAS in improving spatial resolution was verified through simulations. And the superiority of the p-GFMAS united with GA-SVD in enhancing spatiotemporal resolution was validated with in vivo studies. Simulation results demonstrated that p-GFMAS outperforms other compounding methods, and p-GFMAS increased the signal-to-noise ratio (SNR) by 22.77 dB (234%) and reduced the full-width at half-maximum (FWHM) by m (43.4%) over traditional coherent compounding (CC). Rat brain and human liver results depicted that p-GFMAS united with GA-SVD achieved highest image quality when using shorter ensembles due to the increased SNR and reduced FWHM. Dynamic results further revealed the ability of the united method to capture flow-dynamics of the rat brain at a frame rate of 20 Hz. Overall, p-GFMAS united with GA-SVD realizes high spatiotemporal resolution contrast-free microvascular visualization, facilitating an understanding of the mechanisms of various physiological and pathological processes.
engineering, electrical & electronic,instruments & instrumentation
What problem does this paper attempt to address?