Study on the Actuation and Control System of the Guiding Robot for Catheter Based on Heat Transfer Analysis

Liping Xian
Abstract:In order to ensure safety of the operation and prevent overheating of the shape memory alloy(SMA) actuators of the guiding robot used for catheters,we designed an actuation and control system with resistance feedback for the guiding robot.Heat transfer analysis of the SMA actuator was discussed,and the electric current of the SMA actuator driven by pulse-width modulation(PWM) was deduced.Based on heat transfer analysis,a resistance feedback actuation and control system of the guiding robot was developed.In the vascular model,we simulated the catheterization process in the interposition surgery.In light of the scheduled surgery route,the guiding robot entered the target vas ramous successfully under the control of the designed system.The experimental results showed that the invasive surgery process could be simulated accurately and rapidly.The application of the guiding robot can greatly reduces the difficulty of the operation.
Engineering,Medicine
What problem does this paper attempt to address?