Calibration‐free whole‐brain CEST imaging at 7T with parallel transmit pulse design for saturation homogeneity utilizing universal pulses (PUSHUP)
Yannik Völzke,Suzan Akbey,Daniel Löwen,Eberhard Daniel Pracht,Rüdiger Stirnberg,Vincent Gras,Nicolas Boulant,Moritz Zaiss,Tony Stöcker
DOI: https://doi.org/10.1002/mrm.30305
2024-09-21
Magnetic Resonance in Medicine
Abstract:Purpose Chemical exchange saturation transfer (CEST) measurements at ultra‐high field (UHF) suffer from strong saturation inhomogeneity. Retrospective correction of this inhomogeneity is possible to some extent, but requires a time‐consuming repetition of the measurement. Here, we propose a calibration‐free parallel transmit (pTx)‐based saturation scheme that homogenizes the saturation over the imaging volume, which we call PUlse design for Saturation Homogeneity utilizing Universal Pulses (PUSHUP). Theory Magnetization transfer effects depend on the saturation B1rms . PUSHUP homogenizes the saturation B1rms by using multiple saturation pulses with alternating B1 ‐shims. Using a database of B1 maps, universal pulses are calculated that remove the necessity of time‐consuming, subject‐based pulse calculation during the measurement. Methods PUSHUP was combined with a whole‐brain three‐dimensional‐echo planar imaging (3D‐EPI) readout. Two PUSHUP saturation modules were calculated by either applying whole‐brain or cerebellum masks to the database maps. The saturation homogeneity and the group mean CEST amplitudes were calculated for different B1 ‐correction methods and were compared to circular polarized (CP) saturation in five healthy volunteers using an eight‐channel transmit coil at 7 Tesla. Results In contrast to CP saturation, where accurate CEST maps were impossible to obtain in the cerebellum, even with extensive B1 ‐correction, PUSHUP CEST maps were artifact‐free throughout the whole brain. A 1‐point retrospective B1 ‐correction, that does not need repeated measurements, sufficiently removed the effect of residual saturation inhomogeneity. Conclusion The presented method allows for homogeneous whole‐brain CEST imaging at 7 Tesla without the need of a repetition‐based B1 ‐correction or online pulse calculation. With the fast 3D‐EPI readout, whole‐brain CEST imaging with 45 saturation offsets is possible at 1.6 mm resolution in under 4 min.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?