Primary stability and osseointegration comparing a novel tapered design tissue‐level implant with a parallel design tissue‐level implant. An experimental in vivo study

Thomas Gill,Sebastian Kühl,Simon Rawlinson,Benjamin Pippenger,Benjamin Bellon,Shakeel Shahdad
DOI: https://doi.org/10.1111/clr.14301
2024-06-02
Clinical Oral Implants Research
Abstract:Objectives The aim of the present study was to compare a novel tapered, double‐threaded self‐tapping tissue‐Level design implant (TLC) to a well‐established parallel walled tissue‐level (TL) implant in terms of primary and secondary stability over time. Materials and Methods Test TLC (n = 10/per timepoint) and control TL (n = 10/per timepoint) implants were placed in the mandible of minipigs and left for submerged healing for 3, 6, and 12 weeks. Maximum insertion torque and implant stability quotient (ISQ) were measured for each implant at placement. Osseointegration and cortical bone maintenance were histologically evaluated by measuring total bone‐to‐implant contact (BIC) and first bone‐to‐implant contact (fBIC). Results A significantly higher maximum insertion torque was measured for the test implant TLC compared to the control TL implant (57.83 ± 24.73 Ncm and 22.62 ± 23.16 Ncm, respectively; p