An LC–MS-based designated comparison method with similar performance to the Lp(a) reference measurement procedure to guide molar Lp(a) standardization

Nina M. Diederiks,L. Renee Ruhaak,Fred P. H. T. M. Romijn,Mervin M. Pieterse,Nico P. M. Smit,Christa M. Cobbaert
DOI: https://doi.org/10.1186/s12014-023-09446-5
2024-01-26
Clinical Proteomics
Abstract:The 2022 consensus statement of the European Atherosclerosis Society (EAS) on lipoprotein(a) (Lp(a)) recognizes the role of Lp(a) as a relevant genetically determined risk factor and recommends its measurement at least once in an individual's lifetime. It also strongly urges that Lp(a) test results are expressed as apolipoprotein (a) (apo(a)) amount of substance in molar units and no longer in confounded Lp(a) mass units (mg/dL or mg/L). Therefore, IVD manufacturers should transition to molar units. A prerequisite for this transition is the availability of an Lp(a) Reference Measurement Procedure (RMP) that allows unequivocal molecular detection and quantification of apo(a) in Lp(a). To that end an ISO 17511:2020 compliant LC–MS based and IFCC-endorsed RMP has been established that targets proteotypic peptides of apolipoprotein(a) (apo(a)) in Lp(a). The RMP is laborious and requires highly skilled operators. To guide IVD-manufacturers of immunoassay-based Lp(a) test kits in the transition from mass to molar units, a Designated Comparison Method (DCM) has been developed and evaluated.