A remote monitor of bed patient cardiac vibration, respiration and movement

Koji Mukai,Yoshiharu Yonezawa,Hidekuni Ogawa,Hiromichi Maki,W Morton Caldwell
DOI: https://doi.org/10.1109/IEMBS.2009.5333716
Abstract:We have developed a remote system for monitoring heart rate, respiration rate and movement behavior of at-home elderly people who are living alone. The system consists of a 40 kHz ultrasonic transmitter and receiver, linear integrated circuits, a low-power 8-bit single chip microcomputer and an Internet server computer. The 40 kHz ultrasonic transmitter and receiver are installed into a bed mattress. The transmitted signal diffuses into the bed mattress, and the amplitude of the received ultrasonic wave is modulated by the shape of the mattress and parameters such as respiration, cardiac vibration and movement. The modulated ultrasonic signal is received and demodulated by an envelope detection circuit. Low, high and band pass filters separate the respiration, cardiac vibration and movement signals, which are fed into the microcontroller and digitized at a sampling rate of 50 Hz by 8-bit A/D converters. The digitized data are sent to the server computer as a serial signal. This computer stores the data and also creates a graphic chart of the latest hour. The person's family or caregiver can download this chart via the Internet at any time.
What problem does this paper attempt to address?