Dilution with Digital Microfluidic Biochips: How Unbalanced Splits Corrupt Target-Concentration

Sudip Poddar,Robert Wille,Hafizur Rahaman,Bhargab B. Bhattacharya
DOI: https://doi.org/10.48550/arXiv.1901.00353
2019-01-02
Abstract:Sample preparation is an indispensable component of almost all biochemical protocols, and it involves, among others, making dilutions and mixtures of fluids in certain ratios. Recent microfluidic technologies offer suitable platforms for automating dilutions on-chip, and typically on a digital microfluidic biochip (DMFB), a sequence of (1:1) mix-split operations is performed on fluid droplets to achieve the target concentration factor (CF) of a sample. An (1:1) mixing model ideally comprises mixing of two unit-volume droplets followed by a (balanced) splitting into two unit-volume daughter-droplets. However, a major source of error in fluidic operations is due to unbalanced splitting, where two unequal-volume droplets are produced following a split. Such volumetric split-errors occurring in different mix-split steps of the reaction path often cause a significant drift in the target-CF of the sample, the precision of which cannot be compromised in life-critical assays. In order to circumvent this problem, several error-recovery or error-tolerant techniques have been proposed recently for DMFBs. Unfortunately, the impact of such fluidic errors on a target-CF and the dynamics of their behavior have not yet been rigorously analyzed. In this work, we investigate the effect of multiple volumetric split-errors on various target-CFs during sample preparation. We also perform a detailed analysis of the worst-case scenario, i.e., the condition when the error in a target-CF is maximized. This analysis may lead to the development of new techniques for error-tolerant sample preparation with DMFBs without using any sensing operation.
Emerging Technologies
What problem does this paper attempt to address?
The problem that this paper attempts to solve is the error problem of the target concentration (CF, Concentration Factor) caused by unbalanced dispensing operations during sample dilution on digital microfluidic biochips (DMFB). Specifically, the paper focuses on how, during the mixing - dispensing operation, if there are dispensing errors with unequal volumes (that is, the volumes of the two generated sub - droplets are not equal), these errors affect the concentration of the final target sample and may cause the target concentration to deviate significantly from the expected value. Such errors are unacceptable in life - critical detections, so in - depth analysis and research are required to develop new techniques that can tolerate or correct these errors without using any sensing operations. Through theoretical analysis and experimental verification, the paper has verified the influence of multiple volume - dispensing errors on the target concentration, explored the error - maximization conditions in the worst - case scenario, and proposed a method for sample preparation without sensors to ensure that the target concentration is within the allowable error range. This helps to improve the reliability and accuracy of digital microfluidic biochips in the sample preparation process.