In vivo measurement of glycine with short echo-time 1H MRS in human brain at 7 T

Giulio Gambarota,Ralf Mekle,Lijing Xin,Martin Hergt,Wietske van der Zwaag,Gunnar Krueger,Rolf Gruetter
DOI: https://doi.org/10.1007/s10334-008-0152-0
Magma
Abstract:Object: To determine whether glycine can be measured at 7 T in human brain with (1)H magnetic resonance spectroscopy (MRS). Materials and methods: The glycine singlet is overlapped by the larger signal of myo-inositol. Density matrix simulations were performed to determine the TE at which the myo-inositol signal was reduced the most, following a single spin-echo excitation. (1)H MRS was performed on an actively shielded 7 T scanner, in five healthy volunteers. Results: At the TE of 30 ms, the myo-inositol signal intensity was substantially reduced. Quantification using LCModel yielded a glycine-to-creatine ratio of 0.14 +/- 0.01, with a Cramér-Rao lower bound (CRLB) of 7 +/- 1%. Furthermore, quantification of metabolites other than glycine was possible as well, with a CRLB mostly below 10%. Conclusion: It is possible to detect glycine at 7 T in human brain, at the short TE of 30 ms with a single spin-echo excitation scheme.