Scanning Fluorescence Correlation Spectroscopy (SFCS) with a Scan Path Perpendicular to the Membrane Plane

Paul Müller,Petra Schwille,Thomas Weidemann
DOI: https://doi.org/10.1007/978-1-62703-649-8_29
2018-06-01
Abstract:Scanning fluorescence correlation spectroscopy (SFCS) with a scan path perpendicular to the membrane plane was introduced to measure diffusion and interactions of fluorescent components in free standing biomembranes. Using a confocal laser scanning microscope (CLSM) the open detection volume is moved laterally with kHz frequency through the membrane and the photon events are continuously recorded and stored in a file. While the accessory hardware requirements for a conventional CLSM are minimal, data evaluation can pose a bottleneck. The photon events must be assigned to each scan, in which the maximum signal intensities have to be detected, binned, and aligned between the scans, in order to derive the membrane related intensity fluctuations of one spot. Finally, this time-dependent signal must be correlated and evaluated by well known FCS model functions. Here we provide two platform independent, open source software tools (PyScanFCS and PyCorrFit) that allow to perform all of these steps and to establish perpendicular SFCS in its one- or two-focus as well as its single- or dual-colour modality.
Biomolecules,Quantitative Methods
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to measure the diffusion and interaction of fluorescent components in free - standing biofilms. Specifically, the authors introduced a new technique - Scanning Fluorescence Correlation Spectroscopy (SFCS), whose scanning path is perpendicular to the membrane plane. This method can be used to study the molecular mobility and interaction in the lipid bilayer of Giant Unilamellar Vesicles (GUVs). ### Main Problems and Solutions 1. **Problems with Traditional Methods** - **Fixed Detection Volume**: In the measurement process of traditional single - point confocal Fluorescence Correlation Spectroscopy (FCS), the detection volume is fixed in space, and even a tiny fluorescent membrane displacement will cause deviation in the correlation curve. - **Bleaching Problem**: Since the laser focus stays on the membrane for a long time, it will lead to rapid bleaching of fluorescent molecules. 2. **Advantages of SFCS** - **Correcting Position Instability**: SFCS can correct the position instability caused by membrane position changes by determining the maximum intensity on each scanning path through post - measurement. - **Reducing Bleaching**: The laser focus is outside the membrane most of the time, thus reducing the bleaching phenomenon. 3. **Data Processing** - **Software Tools**: The authors provided two open - source software tools (PyScanFCS and PyCorrFit) for processing and analyzing SFCS data. These tools can help extract membrane - related intensity fluctuations and calculate correlation functions. - **Data Calibration**: Calibrate the detection volume through Alexa488 solution to ensure data accuracy. - **Two - focus SFCS**: Determine the distance between parallel scanning paths through the cross - correlation function, so as to measure the diffusion coefficient more accurately. ### Specific Applications - **Single - focus SFCS**: Suitable for single - line scanning, mainly used to measure the diffusion characteristics of single fluorescent molecules. - **Two - focus SFCS**: Through two parallel scanning paths, it can measure the diffusion coefficient more accurately and is suitable for application scenarios requiring high precision. - **Two - color SFCCS**: By alternately exciting fluorescent molecules of different colors, it can study the interaction of different labeled chemicals in biofilms. ### Summary This paper solves the problems of position instability and bleaching in the traditional FCS method when measuring molecular diffusion and interaction in biofilms by introducing the Scanning Fluorescence Correlation Spectroscopy (SFCS) technique perpendicular to the membrane plane. At the same time, the open - source software tools provided make data processing and analysis more convenient and efficient.