Ultra-Low Energy Calibration of LUX Detector using $^{127}$Xe Electron Capture

LUX Collaboration,D. S. Akerib,S. Alsum,H. M. Araújo,X. Bai,A. J. Bailey,J. Balajthy,P. Beltrame,E. P. Bernard,A. Bernstein,T. P. Biesiadzinski,E. M. Boulton,P. Brás,D. Byram,S. B. Cahn,M. C. Carmona-Benitez,C. Chan,A. Currie,J. E. Cutter,T. J. R. Davison,A. Dobi,E. Druszkiewicz,B. N. Edwards,S. R. Fallon,A. Fan,S. Fiorucci,R. J. Gaitskell,J. Genovesi,C. Ghag,M. G. D. Gilchriese,C. R. Hall,M. Hanhardt,S. J. Haselschwardt,S. A. Hertel,D. P. Hogan,M. Horn,D. Q. Huang,C. M. Ignarra,R. G. Jacobsen,W. Ji,K. Kamdin,K. Kazkaz,D. Khaitan,R. Knoche,N. A. Larsen,B. G. Lenardo,K. T. Lesko,A. Lindote,M. I. Lopes,A. Manalaysay,R. L. Mannino,M. F. Marzioni,D. N. McKinsey,D. M. Mei,J. Mock,M. Moongweluwan,J. A. Morad,A. St. J. Murphy,C. Nehrkorn,H. N. Nelson,F. Neves,K. O'Sullivan,K. C. Oliver-Mallory,K. J. Palladino,E. K. Pease,C. Rhyne,S. Shaw,T. A. Shutt,C. Silva,M. Solmaz,V. N. Solovov,P. Sorensen,T. J. Sumner,M. Szydagis,D. J. Taylor,W. C. Taylor,B. P. Tennyson,P. A. Terman,D. R. Tiedt,W. H. To,M. Tripathi,L. Tvrznikova,S. Uvarov,V. Velan,J. R. Verbus,R. C. Webb,J. T. White,T. J. Whitis,M. S. Witherell,F. L. H. Wolfs,J. Xu,K. Yazdani,S. K. Young,C. Zhang
DOI: https://doi.org/10.1103/PhysRevD.96.112011
2017-09-04
Abstract:We report an absolute calibration of the ionization yields($\textit{Q$_y$})$ and fluctuations for electronic recoil events in liquid xenon at discrete energies between 186 eV and 33.2 keV. The average electric field applied across the liquid xenon target is 180 V/cm. The data are obtained using low energy $^{127}$Xe electron capture decay events from the 95.0-day first run from LUX (WS2013) in search of Weakly Interacting Massive Particles (WIMPs). The sequence of gamma-ray and X-ray cascades associated with $^{127}$I de-excitations produces clearly identified 2-vertex events in the LUX detector. We observe the K- (binding energy, 33.2 keV), L- (5.2 keV), M- (1.1 keV), and N- (186 eV) shell cascade events and verify that the relative ratio of observed events for each shell agrees with calculations. The N-shell cascade analysis includes single extracted electron (SE) events and represents the lowest-energy electronic recoil $\textit{in situ}$ measurements that have been explored in liquid xenon.
Instrumentation and Detectors,Instrumentation and Methods for Astrophysics,High Energy Physics - Experiment
What problem does this paper attempt to address?