Highly hydrogen-sensitive thermal desorption spectroscopy system for quantitative analysis of low hydrogen concentration (~1 x 10^16 atoms/cm3) in thin-film samples

Taku Hanna,Hidenori Hiramatsu,Isao Sakaguchi,Hideo Hosono
DOI: https://doi.org/10.1063/1.4982255
2017-05-05
Abstract:We developed a highly hydrogen-sensitive thermal desorption spectroscopy (HHS-TDS) system to detect and quantitatively analyze low hydrogen concentrations in thin films. The system was connected to an in situ sample-transfer chamber system, manipulators, and an rf magnetron sputtering thin-film deposition chamber under an ultra-high-vacuum (UHV) atmosphere of ~10^-8 Pa. The following key requirements were proposed in developing the HHS-TDS: (i) a low hydrogen residual partial pressure, (ii) a low hydrogen exhaust velocity, and (iii) minimization of hydrogen thermal desorption except from the bulk region of the thin films. To satisfy these requirements, appropriate materials and components were selected, and the system was constructed to extract the maximum performance from each component. Consequently, ~2000 times higher sensitivity to hydrogen than that of a commercially available UHV-TDS system was achieved using H+-implanted Si samples. Quantitative analysis of an amorphous oxide semiconductor InGaZnO4 thin film (1 cm x 1 cm x 1um thickness, hydrogen concentration of 4.5 x 10^17 atoms/cm3) was demonstrated using the HHS-TDS system. This concentration level cannot be detected using UHV-TDS or secondary ion mass spectroscopy (SIMS) systems. The hydrogen detection limit of the HHS-TDS system was estimated to be ~1 x 10^16 atoms/cm3, which implies ~2 orders of magnitude higher sensitivity than that of SIMS and resonance nuclear reaction systems (~10^18 atoms/cm3).
Materials Science
What problem does this paper attempt to address?