Technical solutions for simultaneous MEG and SEEG recordings: towards routine clinical use
J M Badier,A S Dubarry,M Gavaret,S Chen,A S Trébuchon,P Marquis,J Régis,F Bartolomei,C G Bénar,R Carron
DOI: https://doi.org/10.1088/1361-6579/aa7655
2017-09-21
Physiological Measurement
Abstract:OBJECTIVE: The simultaneous recording of intracerebral EEG (stereotaxic EEG, SEEG) and magnetoencephalography (MEG) is a promising strategy that provides both local and global views on brain pathological activity. Yet, acquiring simultaneous signals poses difficult technical issues that hamper their use in clinical routine. Our objective was thus to develop a set of solutions for recording a high number of SEEG channels while preserving signal quality.APPROACH: We recorded data in a patient with drug resistant epilepsy during presurgical evaluation. We used dedicated insertion screws and optically insulated amplifiers. We recorded 137 SEEG contacts on 10 depth electrodes (5-15 contacts each) and 248 MEG channels (magnetometers). Signal quality was assessed by comparing the distribution of RMS values in different frequency bands to a reference set of MEG acquisitions.MAIN RESULTS: The quality of signals was excellent for both MEG and SEEG; for MEG, it was comparable to that of MEG signals without concurrent SEEG. Discharges involving several structures on SEEG were visible on MEG, whereas discharges limited in space were not seen at the surface.SIGNIFICANCE: SEEG can now be recorded simultaneously with whole-head MEG in routine. This opens new avenues, both methodologically for understanding signals and improving signal processing methods, and clinically for future combined analyses.
engineering, biomedical,biophysics,physiology