Comparing Equil patch versus traditional catheter insulin pump in type 2 diabetes using continuous glucose monitoring metrics and profiles

Yu‐Jiao Li,Zi‐Yue Shao,Yun‐Qing Zhu,Da‐Shuang Chen,Jian Zhu
DOI: https://doi.org/10.1111/1753-0407.13536
IF: 4.53
2024-04-11
Journal of Diabetes
Abstract:Highlights This study was based on different parameters calculated from continuous glucose monitoring data and demonstrated that the Equil patch insulin pump has similar efficacy in clinical practice as the Medtronic MMT‐712 insulin pump. This study demonstrated that the Equil patch insulin pump had similar efficacy in controlling glucose variability, reducing hyperglycemia, and managing hypoglycemic episodes in type 2 diabetes mellitus patients as the Medtronic MMT‐712 insulin pump. Aims It is not clear whether there are differences in glycemic control between the Equil patch and the MMT‐712 insulin pump. Our objective was to compare two types of insulin pumps in the treatment of type 2 diabetes mellitus (T2DM), using continuous glucose monitoring (CGM) metrics and profiles. Methods This was a randomized case‐crossover clinical trial. Participants were hospitalized and randomly allocated to two groups and underwent two types of insulin pump treatments (group A: Equil patch—Medtronic MMT‐712 insulin pump; group B: Medtronic MMT‐712—Equil patch insulin pump) separated by a 1‐day washout period. Glycemic control was achieved after 7–8 days of insulin pump therapy. Each patient received CGM for 5 consecutive days (from day 1 to day 5). On day 3 of CGM performance, the Equil patch insulin pump treatment was switched to Medtronic MMT‐712 insulin pump treatment at the same basal and bolus insulin doses or vice versa. CGM metrics and profiles including glycemic variability (GV), time in range (TIR, 3.9–10.0 mmol/L), time below range (TBR, 10.0 mmol/L), and postprandial glucose excursions, as well as incidence of hypoglycemia. Results Forty‐six T2DM patients completed the study. There was no significant difference in parameters of daily GV and postprandial glucose excursions between the Equil patch insulin pump treatment and the Medtronic insulin pump treatment. Similarly, there was no between‐treatment difference in TIR, TBR, and TAR, as well as the incidence of hypoglycemia. Conclusion The Equil patch insulin pump was similar to the traditional MMT‐712 insulin pump in terms of glycemic control. Equil patch insulin pump is a reliable tool for glycemic management of diabetes mellitus.
endocrinology & metabolism
What problem does this paper attempt to address?