A high resolution scintillating fiber tracker with SiPM readout
Henning Gast,Thomas Kirn,Gregorio Roper Yearwood,Stefan Schael
DOI: https://doi.org/10.1016/j.nima.2007.08.019
2007-03-15
Abstract:A novel modular high-resolution charged-particle tracking detector design using $250\mu m$ diameter, round, scintillating fibers and SiPM arrays for readout is presented. The fiber modules consist of 8 layers of 128 fibers, mechanically sta bilized by a carbon fiber / Rohacell foam structure. A prototype using scintillating fibers with a diameter of $300\mu m$ and two typ es of silicon photomultipliers has been tested in a $10GeV$ proton beam in late October 2006 at the T9 PS-beamline, CERN. We present the measured spatial resolu tion, efficiency and signal-over-noise for this setup. The advantage of this tra cker concept is a compact and modular design with low material budget and only little or no cooling requirements depending on the quality of the silicon photomultiplieres.
Instrumentation and Detectors