Critical analysis of the relationship between arterial saturation and the ratio-of-ratios used in pulse oximetry

Giles Blaney,Angelo Sassaroli,Sergio Fantini
DOI: https://doi.org/10.1117/1.jbo.29.s3.s33313
IF: 3.758
2024-11-28
Journal of Biomedical Optics
Abstract:SignificancePulse oximetry is a well-established technique for the assessment of arterial oxygen Saturation (SaO2). A quantitative analysis of its basic measurements may benefit its further development.AimWe aim to derive and discuss the quantitative relationship between SaO2 and the measured quantity in pulse oximetry and the ratio-of-ratios (R) of pulsatile intensities at red and infrared wavelengths.ApproachThe source of the pulsatile optical signal is the systolic and diastolic pulsation of the arterial compartment within the probed tissue. From a diffuse optics point of view, this is associated with an absorption change in the region that arterial blood expands into and a mean partial optical path length through that expanded arterial volume. This observation is the basis for a modified Beer–Lambert law analysis of the relationship between SaO2 and R.ResultsThe equation that relates SaO2 and R contains parameters that include the molar extinction coefficients of oxy- and deoxy-hemoglobin and the ratio of the mean partial optical path lengths in the arterial pulsatile expansion volume at the two wavelengths.ConclusionThe analysis reported here clarifies the source of the calibration equations used in pulse oximetry and provides guidance to assess the impact of various key optical parameters on the accuracy of pulse oximetry.
radiology, nuclear medicine & medical imaging,biochemical research methods,optics
What problem does this paper attempt to address?