Development of a 3-Dimensional Dosimetry System for Leksell Gamma Knife-Perfexion

KyoungJun Yoon,JungWon Kwak,DoHeui Lee,ByungChul Cho,SangWook Lee,SeungDo Ahn
DOI: https://doi.org/10.48550/arXiv.1503.03255
2015-03-11
Abstract:The purpose of our study is to develop a new, 3-dimensional dosimetry system to verify the accuracy of dose deliveries in Leksell Gamma Knife-Perfexion TM (LGKP) (Elekta, Norcross, GA, USA). The instrument consists of a moving head phantom, an embedded thin active layer and a CCD camera system and was designed to be mounted to LGKP. As an active material concentrically located in the hemispheric head phantom, we choose Gafchromic EBT3 films and Gd2O2S;Tb phosphor sheets for dosimetric measurements. Also, to compensate the lack of backscatter, we located a 1 cm thick PMMA plate downstream of the active layer. The PMMA plate was transparent for scintillation lights to reach the CCD with 1200x1200 pixels by a 5.2 um pitch. Using this system, three hundred images by a 0.2 mm slice gap were acquired under each of three collimator setups, i.e. 4 mm, 8 mm, and 16 mm, respectively. The 2D projected images taken by CCD camera were compared with the dose distributions measured by EBT3 films in the same conditions. All 2D distributions were normalized by the maximum values derived by fitting peaks for each collimator setup. The FWHM differences of 2D profiles between of CCD images and of film doses were measured to be less than 0.3 mm. The scanning task for whole peak regions took less than three minutes with the new instrument. It could be utilized as a QA tool for the Gamma Knife radiosurgery system instead of film dosimetry, the use of which requires much more time and resources.
Medical Physics
What problem does this paper attempt to address?
This paper aims to develop a new three - dimensional dose measurement system for verifying the dose - delivery accuracy of Leksell Gamma Knife - Perfexion™ (LGKP). Specifically, the research team designed an instrument that includes a moving head phantom, an embedded thin active layer and a CCD camera system, and installed it on the LGKP. The system uses Gafchromic EBT3 film and Gd₂O₂S:Tb phosphor screen as active materials for dose measurement. In addition, to compensate for the lack of backscattering, a 1 - centimeter - thick PMMA plate was placed downstream of the active layer. The PMMA plate is transparent to fluorescent light and can ensure that the light reaches the CCD camera with 1200×1200 pixels and a 5.2 μm pitch. Through this system, the researchers obtained 300 images at a slice interval of 0.2 mm under three different collimator settings (4 mm, 8 mm and 16 mm). These 2D projection images were compared with the EBT3 film measurement results under the same conditions. All 2D distributions were normalized by fitting the peak of each collimator setting. The results showed that the full - width - at - half - maximum (FWHM) difference between the CCD images and the 2D profiles of the film dose was less than 0.3 mm. The scanning task of the entire peak area was completed within three minutes. The study believes that this new device can be used as a quality assurance tool for the gamma - knife radiosurgery system, replacing the traditional film - dose measurement method, which requires more time and resources.