Article original Validation du dosage des espèces de l'arsenic urinaire par HPLC-ICP-MS Arsenic speciation analysis of human urine using HPLC-ICP-MS

É. Saussereau,M. Guerbet,J. Goullé,C. Lacroix
Abstract:Objective: Determination of arsenic species and total arsenic in urine by liquid chromatography (HPLC) and inductively coupled plasma mass spectrometry (ICP-MS) are presented. Methods: The ICP-MS measurements were performed using a Thermo Elemental X7CCT series coupled to a Finnigan Spectro P1000XR HPLC. An anion ex- change Hamilton PRP-X100 column and a gradient elution using ammonium carbonate, allowed the chromatographic separation of the main five arsenic species within 13.5 min: arsenobetaine (AB), arsenite (As III ), dimethylarseniate (DMA V ), monomethylarseniate (MMA V ), arseniate (As V ). Results: Aqueous or urinary calibration produced similar results. The method was linear up to 2000 µg/L total arsenic (i.e. 20 000 µg/L total arsenic for a ten fold urine dilution). Linear regression was higher than 0.998 for the 5 arsenic species. With ten fold urine dilution, the limits of quantitation were: AB = 1.0 µg/L − As III = 1.7 µg/L − DMA V = 3.5 µg/L − MMA = 3.4 µg/L − As V = 1.7 µg/L. Standard addition in control urine charged with 100 µg/L of the species produced recoveries in the range 100.3% (MMA) and 102.8% (DMA). For As III and As V , the most toxic forms of arsenic, the intra-assay and inter-assay variation coefficient were below 10%. Due to the absence of certified reference material containing the 5 species, two commercial samples have been measured. The sum of the five species were 154.6 µg/ L( n = 5), and 114.6 µg/ L( n = 5), for 153 µg/L
Medicine,Environmental Science,Chemistry
What problem does this paper attempt to address?