PCA-Based Preprocessing for Clustering-Based Fetal Heart Rate Extraction in Non-Invasive Fetal Electrocardiograms

Luis Oyarzún,Encarnación Castillo,Luis Parrilla,Uwe Meyer-Baese,Antonio García
DOI: https://doi.org/10.3390/electronics13071264
IF: 2.9
2024-03-29
Electronics
Abstract:Non-invasive fetal electrocardiography (NI-ECG) is based on the acquisition of signals from electrodes on the mother's abdominal surface. This abdominal ECG (aECG) signal consists of the maternal ECG (mECG) along with the fetal ECG (fECG) and other noises and artifacts. These records allow the acquisition of valuable and reliable information that helps ensure fetal well-being during pregnancy. This paper proposes a procedure based on principal component analysis (PCA) to obtain a single-channel master abdominal ECG record that can be used as input to fetal heart rate extraction techniques. The new procedure requires three main processing stages: PCA-based analysis for fECG-component extraction, polarity test, and curve fitting. To show the advantages of the proposal, this PCA-based method has been used as the feeding stage to a previously developed clustering-based method for single-channel aECG fetal heart rate monitoring. The results obtained for a set of real abdominal ECG recordings from annotated public aECG databases, the Abdominal and Direct Fetal ECG Database and the Challenge 2013 Training Set A, show improved efficiency in fetal heart rate extraction and illustrate the benefits derived from the use of such a master abdominal ECG channel. This allows us to achieve proper fetal heart rate monitoring without the need for manual inspection and selection of channels to be processed, while also allowing us to analyze records that would have been discarded otherwise.
engineering, electrical & electronic,computer science, information systems,physics, applied
What problem does this paper attempt to address?
The problem that this paper attempts to solve is the technical challenges encountered in extracting fetal heart rate (fHR) from non - invasive fetal electrocardiogram (NI - ECG). Specifically, the paper focuses on how to effectively extract the fetal electrocardiogram (fECG) signal from abdominal electrocardiogram (aECG) recordings to achieve accurate monitoring of fetal heart rate. aECG recordings usually contain the mother's electrocardiogram (mECG), the fetal electrocardiogram (fECG), as well as other noises and artifacts. These factors make the extraction of fECG signals complicated, especially when the signals are severely interfered or there is a large amount of noise. To overcome these problems, the paper proposes a pre - processing method based on principal component analysis (PCA) to generate a single - channel main abdominal electrocardiogram recording, which can be used as an input for the clustering - based fetal heart rate extraction technique. The main purposes of this method are: 1. **Reduce the influence of noise and artifacts**: Through PCA analysis, the main fECG components can be extracted from multiple aECG channels, thereby reducing the interference of noise and artifacts on the signal. 2. **Improve the accuracy of fHR extraction**: The generated main aECG signal can more accurately reflect the changes in fetal heart rate, thereby improving the efficiency and reliability of fHR extraction. 3. **Avoid manual channel selection**: Traditional fHR extraction methods often require manual selection of aECG channels with better quality, which is not only time - consuming but also depends on the experience of experts. The PCA method can automatically process all channels without manual selection, thereby simplifying the processing flow. 4. **Utilize all available data**: The PCA method can make full use of all aECG channel data. Even if some channels are affected by noise or artifacts, they will not be completely discarded, thereby reducing the loss of information. In conclusion, this paper aims to improve the extraction effect of fetal heart rate in non - invasive fetal electrocardiogram through a new PCA pre - processing method, making it more reliable and efficient in clinical applications.