Analytical performance of a sandwich enzyme immunoassay for pre beta 1-HDL in stabilized plasma

Takashi Miida,Osamu Miyazaki,Yasushi Nakamura,Satoshi Hirayama,Osamu Hanyu,Isamu Fukamachi,Masahiko Okada
DOI: https://doi.org/10.1194/jlr.D200025-JLR200
Abstract:We have established an immunoassay for pre beta 1-HDL (the initial acceptor of cellular cholesterol) using a monoclonal antibody, MAb55201. Because pre beta 1-HDL is unstable during storage, fresh plasma must be used for pre beta 1-HDL measurements. In this study, we describe a method of stabilizing pre beta 1-HDL, and evaluate the analytical performance of the immunoassay for pre beta 1-HDL. Fresh plasma was stored under various conditions with or without a pretreatment consisting of a 21-fold dilution into 50% (v/v) sucrose. Pre beta 1-HDL concentration was measured by immunoassay. In nonpretreated samples, pre beta 1-HDL decreased significantly from the baseline after 6 h at room temperature. Although pre beta 1-HDL was more stable at 0 degrees C than at room temperature, it increased from 30.2 +/- 8.5 (SE) to 56.5 +/- 5.5 mg/l apolipoprotein A-I (apoA-I) (P < 0.001) in hyperlipidemics, and from 18.4 +/- 1.2 to 37.9 +/- 3.3 mg/l apoA-I (P < 0.001) in normolipidemics after 5-day storage. After 30-day storage at -80 degrees C, pre beta 1-HDL increased from 29.0 +/- 4.0 to 38.0 +/- 5.7 mg/l apoA-I (P < 0.001) in hyperlipidemics, whereas it did not change in normolipidemics. In pretreated samples, pre beta 1-HDL concentration did not change significantly under any of the above conditions. Moreover, pre beta 1-HDL concentrations determined by immunoassay correlated with those determined by native two-dimensional gel electrophoresis (n = 24, r = 0.833, P < 0.05). An immunoassay using MAb55201 with pretreated plasma is useful for clinical measurement of pre beta 1-HDL.
What problem does this paper attempt to address?