Validation of image-derived input function using a long axial field of view PET/CT scanner for two different tracers

Xavier Palard-Novello,Denise Visser,Nelleke Tolboom,Charlotte L. C. Smith,Gerben Zwezerijnen,Elsmarieke van de Giessen,Marijke E. den Hollander,Frederik Barkhof,Albert D. Windhorst,Bart NM van Berckel,Ronald Boellaard,Maqsood Yaqub
DOI: https://doi.org/10.1186/s40658-024-00628-0
2024-03-13
EJNMMI Physics
Abstract:Accurate image-derived input function (IDIF) from highly sensitive large axial field of view (LAFOV) PET/CT scanners could avoid the need of invasive blood sampling for kinetic modelling. The aim is to validate the use of IDIF for two kinds of tracers, 3 different IDIF locations and 9 different reconstruction settings.