Characterisation of antibodies and analytes by surface plasmon resonance for the optimisation of a competitive immunoassay based on energy transfer

Ingrid Coille,Günter Gauglitz,Johan Hoebeke
DOI: https://doi.org/10.1007/s00216-001-1179-5
Abstract:The determination of binding constants using surface plasmon resonance (SPR) was introduced to optimise a competitive homogeneous fluorescence energy-transfer immunoassay (ETIA) before labelling. Steroids were chosen as model for the detection of three analytes estrone, estradiol and ethinylestradiol--by taking three polyclonal antibodies (anti estrone-, anti estradiol- and anti estrogen-antibodies) and the corresponding analyte derivatives used for the immunisation. The active concentration of the antibodies was determined before and after labelling. Inhibition curves were recorded using SPR for all possible combinations of analyte, antibody, and analyte derivatives. The experiments revealed that the active antibody concentration can be reduced to 30% whereas the antibody affinity is not affected by the labelling process. Limits of the use of SPR for determination of affinity constants in solution are discussed. All possible ETIA calibration for the quantification of estrone and estradiol was performed. The lower limits of detection for estrone (0.06 microg L(-1)) and estradiol (0.17 microg L(-1)) were reached with the anti-estrogen IgG and its derivative
What problem does this paper attempt to address?