Virtual diluted cone beam CT for device apposition assessment during endovascular treatment of intracranial aneurysm: a technical note

Patricio MUSZYNSKI,Jean François HAK,Basile KERLEROUX,Benjamin GORY,René ANXIONNAT,François ZHU
DOI: https://doi.org/10.1016/j.neurad.2024.01.003
IF: 4.6
2024-01-22
Journal of Neuroradiology
Abstract:The increasing use of embolization devices with suboptimal radio-opacity to treat intracranial aneurysm underscores the need for advanced imaging techniques to characterize device-vessel interactions more accurately. Contrast-diluted cone-beam CT is commonly used in neurointervention but requires additional technical refinements to improve endovascular treatment assessment. In this technical note, we describe the virtual dilution cone beam CT (VDCBCT), a technique that synthetizes non-contrast and contrast-enhanced CBCT images to virtually dilute iodinated contrast agents, thereby facilitating a more accurate assessment of embolization device apposition. Through a set of intracranial aneurysms treated with different embolization devices, we describe the VDCBCT protocol and its usefulness for device apposition confidence. VDCBCT may enhance the global understanding of neurovascular embolization treatments by providing improved visualization of target vessels and low-radio-opacity embolization devices, obviating the need for contrast dilution.
radiology, nuclear medicine & medical imaging,clinical neurology,neuroimaging
What problem does this paper attempt to address?