Suggested guide to using lactate gap as a surrogate marker in the diagnosis of ethylene glycol overdose.

Goce Dimeski,Amanda Holford,Katherine Isoardi
DOI: https://doi.org/10.1177/00045632241292514
2024-10-04
Abstract:BBackground: Ethylene glycol (EG) poisoning, if not diagnosed rapidly, can lead to poor patient outcomes. Gas Chromatography (GC) is primarily used for EG quantitation which is rarely available, and the turn-around time may be prolonged. Most lactate results from point-of care (POCT) methods are falsely elevated in EG poisoning compared with automated chemistry analyser results. In combination the lactate gap (POCT-Automated chemistry) can be used as surrogate marker in just about all laboratories to indicate likely EG toxicity and guide treatment. Case Report: A male presented by ambulance to hospital with severe agitation requiring mechanical ventilation to facilitate ongoing management. Venous blood gas analysis confirmed a high anion gap metabolic acidosis (HAGMA) with an elevated lactate. The lactate and osmolarity measured in the laboratory showed a normal lactate and high osmolarity, giving a large osmolar gap. The patient was immediately commenced on renal replacement therapy for presumed EG poisoning to minimize kidney injury, and the treatment continued for 19 hours. A very high EG concentration was confirmed by GC the next day. Conclusion: An elevated lactate gap along with a HAGMA and osmolar gap can provide rapid surrogate laboratory data indicating EG poisoning enabling timely treatment and better patient outcomes.