Theoretical design of a spherically sectioned phased array for ultrasound surgery of the liver

D R Daum,K Hynynen
DOI: https://doi.org/10.1016/s0929-8266(99)00006-3
Abstract:Goal: The theoretical explanation of the limits of an array transducer to coagulate large tissues volumes. Methods: A theoretical model is used to illustrate the focal limitations of a spherically sectioned array designed for the treatment of deep seated tissue, e.g. liver. The design optimizes the acoustic dose as a function of the focal depth and available acoustic aperture with the goal of coagulating large volumes in a single sonication period. A quantitative measure of the possible region of focal necrosis is modeled as a function of array parameters with the limiting criteria being near field heating and patient pain. Results: Acoustic simulations show that the maximum distance to produce continuous necrosis between foci in a multiple focus pattern and in a temporally multiplexed pattern is approximately 50% larger than the distance needed between sequential foci. Conclusion: Multiple focus patterns or rapidly scanned single foci are significantly advantageous to sequential sonications of a single focus transducer.