On the impact of B0 shimming algorithms on single‐voxel MR spectroscopy

Behrouz Vejdani Afkham,Eva Alonso‐Ortiz
DOI: https://doi.org/10.1002/mrm.30257
2024-08-28
Magnetic Resonance in Medicine
Abstract:Purpose To assess the impact of different B0 shimming algorithms on MRS. Methods B0 field maps and single‐voxel MR spectroscopy were acquired in the prefrontal cortex of five volunteers at 3 T using five different B0 shimming approaches. B0 shimming was achieved using Siemens' proprietary shim algorithm, in addition to the Pseudo‐Inverse (PI), Quadratic Programming (QuadProg), Least Squares (LSq), and Gradient optimization (Grad) algorithms. The standard deviation of the shimmed B0 field, as well as the SNR and FWHM of the measured metabolites, was used to evaluate the performance of each B0 shimming algorithm. Results Compared to Siemens's shim, significant reductions (p